Understanding the limits of ultrasonics for crude oil measurement

Paul Boughton

Raymond J Kalivoda outlines benefits of using ultrasonic flowmeter technology.

Liquid ultrasonic flowmeters (LUFMs) are gaining acceptance in the petroleum industry for a wide range of applications. Initially they were used for non-custody applications. Today, advances in microprocessors, transducers and electronic technology allow multipath LUFMs can provide highly accurate custody transfer flow measurement. They are now accepted and routinely used in many European countries, and API Standard 5.8 Measurement of Liquid Hydrocarbons by Ultrasonic Flowmeters Using Transit Time Technology recognises this technology in North America.

High accuracy and low maintenance are key features that are driving this technology. Ultrasonic meters, like turbine meters are inference meters. They infer the volumetric through-put by measuring the velocity over the flow area. As with all velocity meters, they are Reynolds Number dependent, that is, they are more or less affected by the relationship between velocity and viscosity. They may also be affected by entrained solids, water, gas and wax. These characteristics can affect both the short term accuracy and long term stability of an ultrasonic meter.

Ultrasonic flowmeters for crude oil

Crude oil measurement, unlike refined products, defines a wide range of applications from light condensates with a viscosity of less than 0.5cP to heavy crude oils over 2000cP. The quality of the crude oil, that is the amount and type of containments, also varies widely.
Viscosity can be expressed in many different units. For our purposes kinematic viscosity, which is expressed in centistokes (cSt), is the most suitable. The other commonly used viscosity units in the petroleum industry are: dynamic viscosity; centipoise (cP), which can be converted to centistokes by dividing by the specific gravity,
(cSt = cP / SG); and Saybolt Seconds Universal (SSU) viscosity, which can be changed to centistokes with a conversation chart.
Crude oils are normally defined by their API gravity, which is sometimes confused with the product’s viscosity. API gravity is defined as the density of crude oil at a specific temperature compared to the density of water at a standard temperature of 60°F. The relationship between specific gravity (SG) and API gravity is:

SG (60F/60F) = 141.5 / (131.5 + API)

API gravity is loosely related to viscosity. For light crude oils there is a fairly close relation between viscosity and API gravity. For medium crude oils and heavy crude oils, it is important to obtain the viscosity from the assay or from a specific viscosity test.

The viscosity of all liquids varies with temperature. The effect of temperature for medium and heavy crude oils can significantly change a meter’s performance due to the considerable change in viscosity. For this reason, it is important when evaluating any meter application that the viscosity of each product must be specified over the operating temperature range.

Fluid properties

Sediment and Water (S&W) is a collective term for non-hydrocarbons found in crude oil. In API MPMS Chapter 1, S&W is defined as: “A material, coexisting with yet foreign to a petroleum liquid ... may include free water and sediment (FW&S) and emulsified or suspended water and sediment (SW&S).”

Since all pipelines regulate the amount of S&W they will accept, normally less than 1 per cent, a crude oil within these requirements is termed ‘pipeline quality oil’. In general, the free water in ‘pipeline quality oil’ should not be a problem for measurement with ultrasonic meters. Sediment, though, especially small particulates may present a problem because they can diffuse the ultrasonic signal.

Gas slugs or entrained gas will not damage an ultrasonic meter but can adversely affect the measurement accuracy of the meter.  Even a small number of gas bubbles can cause attenuation of the ultrasonic signal. The degree of attenuation depends on a number of factors, such as pressure, bubble size, amount of free gas, temperature and signal frequency.

Wax crystals begin to form in a petroleum product at a specific temperature known as the cloud point. If a meter is operated below the cloud point, wax can form on the measurement element, which can notably affect the meter’s accuracy. Some meters are considerably more tolerant of waxing than other meters. For example, after an initial build-up of wax on the walls of positive displacement meter, rotating blades wipe the surfaces, and the meter factor remains stable. In the case of velocity meters (eg, ultrasonic, turbine) there is a continuous meter factor shift as the wax builds up.

Principle of operation

The operating principal of ultrasonic meters is shown in (Fig 1). The volume throughput (Q) is equal to the fluid velocity measured (Vm) multiplied by the area (A) or Q = Vm x A, where the fluid velocity measured is proportional to the difference between an ultrasonic signal traveling with the flow (tAB) and against the flow (tBA). The measurement principal is fairly simple but there are number of factors that must be addressed to achieve custody transfer measurement accuracy.

A misconception exists that ultrasonic meters are not sensitive to fluid properties. This is not the case. To achieve the level of precision measurement available with other metering technologies, these affects must be addressed. This is especially important with crude oil measurement as the oil is typically highly viscous with high levels of contaminants.

On a qualitative level the influences of fluid properties have been addressed by various authors. Knowledge on the quantitative affects of fluid properties on ultrasonic meter accuracy is limited.
The influence of fluid properties on the UFM performance may be classified in two main groups:

●  Signal quality affects – the signal attenuation and signal-to-noise ratio (SNR) in the acoustic paths
●  Flow profile affects – the robustness of the integration method used to combine the individual acoustic path measurements into a full volumetric flow rate measurement

The signal quality of an ultrasonic meter in crude oil applications is determined by: viscosity, entrained gas, sediment and water (S&W) and wax content.
The signal strength, or more precisely, the signal-to-noise ratio (SNR), is crucial for the accuracy of the transit time measurements made in the LUFM. Reduced SNR can mean higher uncertainty of the transit time measurement, resulting in a higher uncertainty of the volumetric flow rate measurement.  In the worst case, the signal can’t be discerned from the noise and the measured output is erroneous.
Noise is classified as:

●  Coherent noise (signal interference) which includes:
a. Transducer ‘ringing’ effects;
b. Spool-piece borne signals (acoustic cross talk); c. Liquid borne reflections (transducer ports reflections, pipe wall reflections/reverberation).

●  Incoherent noise (‘signals’ with random phase relative to the measurement signal) includes:
a.electromagnetic noise (RFI);
b. flow noise;
c. valve noise;
d. structural (pipe) vibrations, etc.

The ‘strength’ (amplitude) of the measurement signal has to ‘compete’ with noise, to give a sufficient SNR. A number of fluid-dependent factors can attenuate the measurement signal and decrease the SNR. The factors that contribute to the sound attenuation coefficient, α, are:

α = αabs + αwio + αgas + αsolids + αwax,

Where:

αabs is the sound absorption coefficient of the pure fluid, and
αwio + αgas + αsolids and αwax account for contaminants in the crude oil that cause excess attenuation due to entrained water, gas, solid particles, and (wax) at the transducers.

Sound absorption (αabs) is the attenuation due to viscosity and the influence of absorption on the SNR. The absorption increases with viscosity and the distance between transducers, therefore the larger the meter diameter the higher the absorption.

Water droplets (αwio) in the oil cause excess sound attenuation due to scattering of the sound waves by the droplets. The effect on the SNR is accounted for in the attenuation coefficient αwio. The coefficient is determined by the water droplet size and distribution, the amount of water in the oil, the pressure and temperature, the oil type and the ultrasonic flowmeter signal frequency. Because of the complexity of this relationship, it is difficult to determine the attenuation coefficient with a high degree of certainty. In the development phase of one recently introduced liquid ultrasonic meter, extensive testing was done on water-in-oil affects on signal attenuation. The following is a summary of these tests:

●  The ultrasonic meter may operate with up to
5 per cent water-in-oil depending on the size of the water droplets.
●  The influence of pressure on water-in-oil absorption is minimal as expected because of the low compressibility of the fluids.
●  Temperate can heighten or lower the attenuation depending on a number of factors.  

Free gas-in-oil (αgas) in the form of gas bubbles, causes excess sound attenuation due to scattering of the sound waves by the bubbles and bubble resonances. The parameters that affect this coefficient are: bubble size and distribution, the amount of free gas present in the oil, the pressure and temperature, the oil type and the LUFM operating frequency. Gas-in oil is a highly complex condition that can have a profound affect on performance. As found in the development of the liquid ultrasonic meter cited above, even small amounts of entrained gas can adversely affect ultrasonic meter performance. A summary of the gas-in-oil affects on signal attenuation are:

●  For small bubbles (less than 0.5 mm diameter) a concentration as low as 1,000 ppm
(0. 1 per cent) can momentarily or completely interrupt the measurement signal.
●  For bubbles over 0.5 mm diameter a concentration up to 10,000 ppm (1%) may be tolerable before the SNR is reduced to a critical level.
●  The attenuation is based on path length so the larger diameter meters are proportionately more affected by entrained gas than smaller diameter meters.
●  Low pressure can significantly increase the signal attention.
●  Temperature has minimal affect.

Solid particles-in-oil (αsolids) like free gas-in-oil, can cause excess sound attenuation due to scattering of the sound waves by the particles. The same parameters that affect free gas-in-oil also affect the αsolids coefficient. The effect of solid particles-in-oil has not been studied in detail. Cursory testing shows results similar to gas-in-oil.

Wax can affect the SNR and the meter’s K-factor. If the temperature is below the cloud point, wax contamination may build up at different surfaces of the ultrasonic flowmeter. Possible influences which may be important for the ultrasonic flowmeter’s performance include the following:

Wax layer build-up:

●  At the transducer, fronts may shift the transit times and cause a continuous meter factor shift as the wax builds up. Cause attenuation (αwax) can reduce the SNR. Due to the relatively small difference in the acoustic impedance between oil and wax, a thin wax layer may not affect the SNR significantly, unless the layer becomes thick, and not homogeneous.  
●  In the transducer, cavities may reduce the acoustic isolation of the transducer from the spool piece, and increased acoustic “cross-talk” through the spool piece. Since cross-talk acts as coherent noise, this results in reduced SNR, and thus can reduce accuracy of the transit time measurements.
●  A wax layer build-up on the inner wall of the spool piece reduces the flow area, which increases the meter’s K-factor. Since flow area A is equal to the pipe radius squared (A = πR2) even a relatively small build-up can cause a large measurement error. For example, a 0.1 mm build-up on the interior diameter of a six-inch meter will result in a 0.27 per cent measurement error. For a 20-inch meter, the error is proportionally smaller, about 0.08 per cent. Regular in-situ proving of the flow meter will correct for misreading due to such wax build-up.

Proving recommendations

In-situ proving at regular intervals is recommended to maintain optimum measurement accuracy.

Ultrasonic meters, as previously stated, are like turbine meters in that they infer the volumetric throughput by measuring the velocity over the flow area. For low viscosity products the velocity profile is flat and the flow velocity is nearly constant over the flow area, except for a region near the pipe wall (Fig.2b). Therefore, the average stream velocity can be measured at any point except near the pipe wall. As the viscosity increases and/or the flow decreases, the flow profile becomes parabolic (Fig. 2a). Maximum velocity is at the center of the pipe and the velocity decreases gradually to zero at the pipe wall. To determine the average stream velocity for this type of profile, the stream velocity is measured at several selective points and the velocities are integrated with an algorithm to determine the average velocity. The relationship between velocity and viscosity is defined by Reynolds Number which is the ratio of flow rate to the meter size and the viscosity {Re No ≈ (flow rate)/(meter size x viscosity).

A select few multi-path ultrasonic meters use velocity from four chordal paths (Fig. 3) with the VPC to accurately determine the average velocity over the complete flow and viscosity range.

Metering systems can also have valves, strainers, elbows, tees, and headers upstream of the meter.  These elements can distort the flow profile and introduce swirl and crossflow upstream of the meter. Since we are measuring velocity, any change created by these elements will affect the measurement accuracy. Flow conditions are used to minimize these affects but a robust integration method with crossflow compensation is also important to optimize performance At least one liquid ultrasonic meter utilises two additional paths with the VPC to correct for swirl and cross flow (Fig. 4).

Specifying an ultrasonic flowmeter

Key characteristics to look for in an ultrasonic meter for crude oil include:

●  A multipath meter with an integration method, like VPC, to improve performance on high-viscosity low-Reynolds Numbers applications.
●  Robustness in correcting asymmetric axial flow velocity profiles.
●  Compensation transverse (non-axial) flow components (swirl, crossflow, etc).

The key benefits of Ultrasonic Technology, low pressure and low maintenance are highly attractive for crude oil measurement. They are driving the technology, but as with any meter, there are limitations. Like turbine meters, they are best operated at the higher flow ranges for optimum accuracy. However, with techniques such as VPC, accurate measurement at the lower flow ranges can also be achieved. No pressure loss results in reduced operating costs. No moving parts results in increased service life and may reduce the frequency of proving since usage wear is a key reason why meters must be recalibrated. 

The measurement technique is susceptible to installation effects and fluid properties. As with all meters, liquid ultrasonic flowmeters need to be proven. In-situ proving, though difficult, is indisputably the best method to reduce the total measurement uncertainty. Proving the meters in a laboratory offers an alternative, but at substantially higher risk of measurement error. Even though a specific ultrasonic meter may compensate for installation effects, such as swirl or cross flow, there isn’t any means of verifying this without field proving. The measurement accuracy is further compromised by the fluid property effects that were discussed. This is especially true for crude oils, because their properties are difficult to simulate even in a laboratory that tests with hydrocarbon fluids. Until more data is available quantifying how liquid ultrasonic flowmeters react to different installation conditions, arguments advocating methods other than in-situ proving under actual operating conditions are questionable.

Ultrasonic meters can provide accurate measurement over a wide range of crude oil applications if they are properly applied, proven and operated.●

Enter √ or  at www.engineerlive.com/iog

Raymond J. Kalivoda is petroleum measurement development manager for FMC Technologies, Erie, PA, USA. www.fmctechnologies.com

Fig. 1. Operating principal of liquid ultrasonic flowmeters.
 
Fig. 2a. High viscosity. Fig. 2b. Low viscosity. 
 
Fig. 3. Four-path arrangements for velocity profile correction. 
 
Fig. 4 .Two added paths for swirl and crossflow correction.